February 12, 2015
by Alexander Gaffney, RAC

Citing Safety Risks, FDA Adopts New Standards for Small-Bore Connectors

Even small devices can cause big problems. That's the conclusion of a new guidance document issued this week by the US Food and Drug Administration (FDA), which is trying to make "small-bore connectors" safer.

Background

The devices—technically a medical device—are used to temporarily connect two otherwise incompatible medical devices, FDA explains in its final guidance document, Safety Considerations to Mitigate the Risks of Misconnections with Smallbore Connectors Intended for Enteral Applications.

For example, a small-bore connector might connect an IV tube implanted in a patient's arm to a bag of saline solution, or a feeding tube implanted in a patient's stomach to enteral nutrition.

The problem, FDA explains in the guidance, is that the "use of common connector designs, such as Luer connectors, has led to unintended connections between devices that have different intended uses." This, FDA added, has had fatal consequences for some patients.

Device connectors have been on FDA's radar for other, related reasons in recent years. For example, in April 2013 the agency released a guidance establishing a new standard intended to ensure that pre-filled glass syringes would be able to connect to receiving devices like Luer locks. Regulators recounted that incompatibility between the products had in some cases delayed treatment of patients in life-threatening situations.

A New Standard for Enteral Devices

FDA's latest guidance, though, confronts the opposite of that problem—devices are too easy to connect together, which has resulted in some patients receiving the incorrect treatment to an improper area.

Now regulators have put the device industry on notice, saying that they’re toughening standards meant to prevent a mismatched connection. "Color-coding, labeling and tagging attached to the device, by themselves, are no longer sufficient to satisfy safety concerns regarding misconnection hazards associated with connectors," FDA explains in the guidance.

Instead, all devices—including small-bore connectors already on the market—will need to comply with new standards developed by the Association for the Advancement of Medical Instrumentation (AAMI) and adopted this week by FDA.

The standards, AAMI/CN3:2014 (PS) Part 3 and AAMI/CN20(PS):2014, aren't available online, but a synopsis of the standards on AAMI's website indicates that the standards call for different physical designs for each connector, and the abandonment of a "universal design."

FDA said devices requiring modification to meet the new standard will be able to submit a declaration of conformity to the standard to modify their existing 510(k) application. This policy will allow companies to update their devices without needing to file a new 510(k), FDA said.

 

Safety Considerations to Mitigate the Risks of Misconnections with Smallbore Connectors Intended for Enteral Applications (FR)

×

Welcome to the new RAPS Digital Experience

We have completed our migration to a new platform and are pleased to introduce the updated site.

What to expect: If you have an existing login, please RESET YOUR PASSWORD before signing in. After you log in for the first time, you will be prompted to confirm your profile preferences, which will be used to personalize content.

We encourage you to explore the new website and visit your updated My RAPS page. If you need assistance, please review our FAQ page.

We welcome your feedback. Please let us know how we can continue to improve your experience.