The European Commission is offering guidance for device companies and other stakeholders trying to determine if their standalone software should be considered a medical device, in vitro diagnostic (IVD) or neither.
As the commission notes, standalone software can directly control an apparatus (e.g. radiotherapy treatment), provide immediate decision-triggering information (e.g. blood glucose meters) or provide support for healthcare professionals (e.g. ECG interpretation), though not all such software programs can be regulated as a medical device or IVD.
The guidance outlines six steps to make such a determination for devices:
In addition, standalone software fulfilling the definition of medical device and intended to be used for the purpose of providing information derived from in vitro examination of a specimen derived from the human body falls under Directive 98/79/EC.
Provided that standalone software is intended specifically by its manufacturer to be used together with an IVD medical device to enable that device to be used in accordance with its intended purpose, this standalone falls under the scope of the IVD Directive and shall be treated as an IVD device in its own right (ie. analysis and interpretation of the optical density delivered by an ELISA reader, line or spot pattern of a blot).
For IVDs, the guidance says that, “If the information provided by the software is based on data obtained from IVD medical devices only, the software is an IVD medical device or an accessory of an IVD medical device. If data are obtained from both IVD medical devices and from medical devices are analysed together for the purpose of providing information according to the definition of an IVD medical device, this software is an IVD medical device (e.g. evaluation of the risk of Trisomy).”
For standalone software intended for use in combination with other devices or equipment, the whole combination must be safe and must not impair the specified performances of the devices, the commission says.
In the guidance, the commission offers several examples of what types of standalone software should and should not be considered devices or IVDs.
For instance, hospital information systems (systems that support the process of patient management and are intended for patient admission, for scheduling patient appointments, for insurance and billing purposes) are not qualified as medical devices.
But if they are computer-based tools combining medical knowledge databases and algorithms with patient-specific data and are intended to provide healthcare professionals and/or users with recommendations for diagnosis, prognosis, monitoring and treatment of individual patients, then they would be considered medical devices.
For example, radiotherapy treatment planning systems, which are intended to calculate the dosage of ionizing irradiation to be applied to a specific patient, are considered to control, monitor or directly influence the source of ionizing radiation and qualify as medical devices, according to the commission.
In addition, drug (e.g.: chemotherapy) planning systems intending to calculate the drug dosage to be administered to a specific patient also qualify as medical devices.
Clinical information systems (CIS) or patient data management systems (PDMS) that store and transfer patient information generated in association with the patient’s intensive care treatment (and usually containing patient identification, vital intensive care parameters and other documented clinical observations) do not qualify as medical devices. But modules that are intended to provide additional information contributing to diagnosis, therapy and follow-up (e.g. generate alarms) do qualify as medical devices.
Another example from the commission is a system for managing pre-hospital electrocardiographs (ECGs), which is a software-based system intended for ambulance services to store and transfer information from patients connected to an ECG monitor to a doctor at a remote location.
“Usually the system contains information about patient identification, vital parameters and other documented clinical observations,” the commission says, noting that these systems do not qualify as medical devices. But if there are modules that provide a patient's treatment information to the paramedics in the ambulance to start the patient’s treatment while the patient is being transported, then the software would qualify as a medical device.
We have completed our migration to a new platform and are pleased to introduce the updated site.
What to expect: If you have an existing login, please RESET YOUR PASSWORD before signing in. After you log in for the first time, you will be prompted to confirm your profile preferences, which will be used to personalize content.
We encourage you to explore the new website and visit your updated My RAPS page. If you need assistance, please review our FAQ page.
We welcome your feedback. Please let us know how we can continue to improve your experience.