FDA Warns Abbott Device Manufacturing Plant in California

The warning letter sent Wednesday from the US Food and Drug Administration (FDA) details four observations on failures related to corrective and preventive actions (CAPAs), controls, design verification and design validation.

The warning letter for the Sylmar, CA-based site, which manufactures implantable cardioverter defibrillators and cardiac resynchronization therapy defibrillators, followed a 10-day FDA inspection in February. FDA previously issued a Form 483 for the site in February.

Abbott told Focus in a statement: "At Abbott, patient safety comes first. We have a strong history and commitment to product safety and quality, as demonstrated by our operations across the company. Abbott acquired St. Jude Medical in January 2017; the FDA inspection of the Sylmar facility, formerly run by St. Jude Medical, began on February 7; and we responded to the 483 observations on March 13, describing the corrective actions we are implementing. We take these matters seriously, continue to make progress on our corrective actions, will closely review FDA's warning letter, and are committed to fully addressing FDA's concerns." 

CAPAs

In one instance, FDA found that Abbott “underestimated the occurrence” of a possibly hazardous situation involving premature battery drainage.

FDA also found the company did not “confirm that verification or validation activities for the corrective actions had been completed, to ensure the corrective actions were effective and did not adversely affect the finished device.”

Controls

On 11 October 2016, Abbott initiated a recall for its Fortify, Unify and Assura Implantable Cardioverter Defibrillators (ICDs) and Cardiac Resynchronization Therapy Defibrillators (CRT-Ds) due to premature battery depletion. 

But FDA found that subsequently, 10 ICDs subject to this recall were shipped from distribution centers to St. Jude US field representatives. And between 14 and 26 October, an additional seven ICDs, also subject to the same recall and in the control of St. Jude US field representatives, were implanted into patients.  

Design Verification and Validation

In addition, FDA found Abbott failed to ensure that “design verification shall confirm that the design output meets the design input requirements,” and failed to accurately incorporate the findings of a third-party assessment into updated cybersecurity risk assessments for high voltage and peripheral devices.

Warning Letter

×

Welcome to the new RAPS Digital Experience

We have completed our migration to a new platform and are pleased to introduce the updated site.

What to expect: If you have an existing login, please RESET YOUR PASSWORD before signing in. After you log in for the first time, you will be prompted to confirm your profile preferences, which will be used to personalize content.

We encourage you to explore the new website and visit your updated My RAPS page. If you need assistance, please review our FAQ page.

We welcome your feedback. Please let us know how we can continue to improve your experience.