| Regulatory Citation | Short Description | Long Description | Frequency |
|---|
| 21 CFR 820.100(a) | Lack of or inadequate procedures | Procedures for corrective and preventive action have not been [adequately] established. | 360 |
| 21 CFR 820.198(a) | Lack of or inadequate complaint procedures | Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been [adequately] established. Specifically,*** | 251 |
| 21 CFR 820.50 | Purchasing controls, Lack of or inadequate procedures | Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been [adequately] established. | 129 |
| 21 CFR 820.75(a) | Lack of or inadequate process validation | A process whose results cannot be fully verified by subsequent inspection and test has not been [adequately] validated according to established procedures. | 122 |
| 21 CFR 803.17 | Lack of Written MDR Procedures | Written MDR procedures have not been [developed] [maintained] [implemented]. | 117 |
| 21 CFR 820.100(b) | Documentation | Corrective and preventive action activities and/or results have not been [adequately] documented. | 101 |
| 21 CFR 820.90(a) | Nonconforming product, Lack of or inadequate procedures | Procedures have not been [adequately] established to control product that does not conform to specified requirements. | 100 |
| 21 CFR 820.30(i) | Design changes - Lack of or Inadequate Procedures | Procedures for design change have not been [adequately] established. Specifically,*** | 95 |
| 21 CFR 820.22 | Quality audits - Lack of or inadequate procedures | Procedures for quality audits have not been [adequately] established. | 90 |
| 21 CFR 820.198(c) | Investigation of device failures | Complaints involving the possible failure of [a device] [labeling] [packaging] to meet any of its specifications were not [reviewed] [evaluated] [investigated] where necessary. | 68 |
| 21 CFR 820.25(b) | Training - Lack of or inadequate procedures | Procedures for training and identifying training needs have not been [adequately] established. | 61 |
| 21 CFR 820.40 | Procedures not adequately established or maintained | Document control procedures have not been adequately [established] [maintained]. Specifically,*** | 61 |
| 21 CFR 820.20(c) | Management review - Lack of or inadequate procedures | Procedures for management review have not been [adequately] established. Specifically,*** | 60 |
| 21 CFR 820.181 | DMR - not or inadequately maintained | A device master record has not been [adequately] maintained. | 58 |
| 21 CFR 820.80(d) | Lack of or inadequate final acceptance procedures | Procedures for finished device acceptance have not been [adequately] established. | 55 |