Industry groups want the US Food and Drug Administration (FDA) to clarify the scope of a recently published guidance on assessing potency assays for active immunotherapy products (ACTIMP) and ensure it aligns with a recently finalized broader cell and gene therapy (CGT) guidance. A patient advocacy group also asked the agency to clarify its expectations regarding the use of non-animal assays to determine the potency of ACTIMPs.
In August, the FDA published a draft guidance intended to help sponsors assess potency assays for a subset of CGT products called ACTIMPs. The guidance covers products such as certain cancer vaccines and treatments for autoimmune diseases and is intended to supplement another guidance document titled “Potency Assurance for Cellular and Gene Therapy Products,” which was finalized earlier this month (RELATED: FDA drafts guidance on assessing potency of immunotherapy products, Regulatory Focus 23 August 2026 and FDA finalizes cell and gene therapy potency guidance, Regulatory Focus 2 November 2026).
The CGT industry group, the Alliance for Regenerative Medicine (ARM), commented on the draft guidance and asked FDA to ensure it aligns with the final potency assurance guidance. More specifically, it asked the agency to update all references to the final potency assurance guidance and explicitly state that the ACTIMP guidance will not increase requirements for sponsors.
"The ACTIMP-specific recommendations should operationalize, rather than narrow, the science- and risk-based flexibility established in the final Potency Assurance guidance," said ARM. "FDA should also clarify that the product-type recommendations in Section V are illustrative applications of a risk-based potency assurance strategy, not additional expectations.
"Several provisions of the final [potency assurance] guidance are directly relevant to ACTIMPs and should be carried forward or cross-referenced," the group added.
Potency strategies
While ARM agrees with FDA's iterative approach to potency assurance and assay development, it is concerned that some of the draft recommendations may be interpreted as fixed expectations regardless of the product development stage. The group asked the agency to use phase-appropriate language throughout the guidance, consistent with the International Council for Harmonization (ICH) principles and the potency assurance guidance, and to provide examples of acceptable potency strategies that sponsors may consider at different stages of product development.
ARM asked FDA to clarify that while it states that quantitative bioassay is the most direct way to assess ACTIMP potency, it is not intended to create a hierarchy of scientifically justified approaches. Instead, the group said that appropriate approaches to understanding the product's potency should be based on the product itself, its mechanism of action (MOA), development stage, and the totality of the potency assurance strategy.
"Consistent with ICH Q6B, it is not necessary for a potency assay to reproduce the full biological cascade; it should measure biological activity relevant to the product’s intended MOA," said ARM. "Assays measuring attributes with an established relationship to biological function (e.g., activation marker expression following stimulation), engineered reporter cell lines, and cell-based expression assays for vectored products may be appropriate when scientifically justified."
Biocom, a life sciences group based in California, asked FDA to clarify what it means for a potency assay to be part of an effective potency assurance strategy and include language in the introduction describing key elements of an ACTIMP potency assurance strategy. More specifically, it asked the agency to explicitly state that an effective potency assurance strategy does not necessarily require a single assay to reproduce the entire MOA or predict clinical efficacy. Furthermore, it asked the agency to clarify that neither the individual potency assay nor the broader potency assurance strategy should be considered an assessment of product design safety.
"If the Agency intends to address potency assurance strategies in a separate guidance document, FDA should cross-reference that guidance to provide sponsors with a clear understanding of the broader framework within which potency assays are developed and evaluated," said Biocom. "Such clarification on these foundational concepts would improve transparency, promote consistent implementation across sponsors and review divisions, and help ensure that potency assurance strategies are developed in a manner that is scientifically appropriate, risk-based, and aligned with FDA’s stated objectives for the guidance."
Biocom asked FDA to clarify that the strength and type of evidence to identify potency-related critical quality attributes (CQAs) should be phase-appropriate because the relationship between the MOA and CQAs may continue to evolve as product development progresses and more is learned about the MOA. It also asked the agency to clarify that acceptance criteria should be phase-appropriate, quantitatively justified, capable of rejecting sub-potent lots, and progressively refined using manufacturing, nonclinical, and clinical data.
Furthermore, Biocom asked for more illustrative examples and relevant scientific references describing assay types, so sponsors can better understand acceptable direct measures of potency. It also asked for more clarity on how to determine what information is sufficient to ensure adequate potency when a physicochemical assay alone is used to measure potency.
Scope
Another key issue ARM wants FDA to address is clarifying the scope of the ACTIMP guidance. The group is concerned that the draft guidance's description of cell-based ACTIMPs as products used to modulate immune responses could be interpreted to encompass immunomodulatory cell therapies that do not act through an antigen, which it argued should be outside the scope of the guidance. To address the issue, it asked that the agency confirm that the guidance applies only to products whose MOA depends on the delivery of a defined or undefined antigen, and that potency assurance for other CGTs is addressed in the final Potency Assurance guidance. Furthermore, it asked regulators to confirm that conventional monoclonal antibody therapeutics that do not function as active immunotherapies are also outside the guidance's scope.
As currently written, ARM also noted that the qualification of bioinformatics pipelines and the assessment of pipeline changes are outside the scope of the guidance, even though they may affect predicted epitopes and are subject to reporting requirements.
"ARM recommends that FDA address expectations for pipeline qualification, justification, and lifecycle management, either in the final guidance or in a companion document, including risk-based considerations for assessing and managing pipeline changes and the supporting data that may be appropriate," said the group.
Biocom also asked FDA to clarify the scope of the guidance. It argued that some of the language in the guidance may create uncertainty regarding whether potency assays are expected to assess product-design safety in addition to intended biological activity.
"We suggest that FDA more clearly distinguish risk-informed selection and control of potency attributes from the assessment of safety hazards," said Biocom. "Potency assays are intended to demonstrate that a product retains its intended biological activity and should not be used to evaluate unintended immune effects or establish clinical safety."
Non-animal-based assays
The patient advocacy group, the Cure Coalition, also commented on the draft guidance, expressing concern about the reliability of potency testing and the conditions under which a non-animal method may replace an animal-based assay. The group notes that the draft guidance doesn't elaborate on the validation requirements for a sponsor proposing to replace an animal-based assay with a non-animal-based assay.
"In this instance, we understand that validating a potency assay generally addresses whether the assay performs reliably," said the Coalition. "A replacement must also show that it measures the same biological activity that the animal-based assay was used to measure.
Because potency testing is applied to lots across a product’s life, a substitute that is reliable but measures the wrong thing would create a recurring gap in quality assurance," the group added.
Ultimately, the Coalition asked FDA to make changes stating that the replacement of animal-based assays is encouraged when a non-animal method has been validated for the specific potency measurement it would replace, and that such validation should be able to measure the biological activity relevant to the product’s MOA.
"The Cure Coalition supports the FDA’s effort to provide clear potency assessment expectations, and we support the 3Rs principles referenced in the guidance," said the group. "We submit these comments and recommendations because we believe that encouraging replacement of animal-based potency assays without a defined validation condition does not provide adequate assurance that a substitute measures what it is intended to measure."